Acute obstruction by pannus in patients with aortic medtronic-hall valves: 30 years of experience

VS Ellensen, KS Andersen, N Vitale… - The Annals of Thoracic …, 2013 - Elsevier
VS Ellensen, KS Andersen, N Vitale, ES Davidsen, L Segadal, R Haaverstad
The Annals of Thoracic Surgery, 2013Elsevier
Background Acute dysfunction of mechanical aortic valve prostheses is a life-threatening
adverse event. Pannus overgrowth, which is fibroelastic hyperplasia originating from the
periannular area, is one cause of dysfunction. The aim of this study was to determine the
annual incidence of readmittance resulting from acute obstruction caused by pannus during
30 years of observation in patients with Medtronic-Hall aortic valve prostheses and to
analyze the risk factors associated with pannus development. Methods From 1982 to 2004 …
Background
Acute dysfunction of mechanical aortic valve prostheses is a life-threatening adverse event. Pannus overgrowth, which is fibroelastic hyperplasia originating from the periannular area, is one cause of dysfunction. The aim of this study was to determine the annual incidence of readmittance resulting from acute obstruction caused by pannus during 30 years of observation in patients with Medtronic-Hall aortic valve prostheses and to analyze the risk factors associated with pannus development.
Methods
From 1982 to 2004, 1,187 patients in our department underwent aortic valve replacement with Medtronic-Hall mechanical monoleaflet valve prostheses. As of December 31, 2012, 27 of these patients (2.3%) had presented with acute valve dysfunction caused by pannus obstruction.
Results
The annual incidence of pannus was 0.7 per 1,000. The median time from the primary operation to prosthetic dysfunction was 11.1 years (range, 1.2 to 26.8 years). Of the 20 patients who underwent reoperation, 2 died. Seven patients died before reoperation. Women had a higher risk for the development of obstructing pannus, and patients with pannus obstruction were younger. Valve size was not an independent risk factor.
Conclusions
Women and younger patients are at higher risk for pannus development. When acute dysfunction by pannus is suspected in a mechanical aortic valve, an immediate echocardiogram and an emergency aortic valve replacement should be carried out because of the potential of a fatal outcome.